Chapter 8 Atomic absorption spectrometry (AAS)
Chapter 8 Atomic absorption spectrometry (AAS)
●Basic principle of atomic absorption spectrometric analysis ● Structure of atomic absorption spectrophotometer ●Interference and elimination ●Quantitative analytical methods ●Sensitivity, characteristic concentration and detected limit ●brief introduction of atomic fluorescence spectrometry (AFS) 掌握原子吸收光谱分析法的基本原理,了解原子吸收光谱 仪的构造,重点掌握原子吸收分光光度定量分析方法。 Outlines
●Basic principle of atomic absorption spectrometric analysis ● Structure of atomic absorption spectrophotometer ●Interference and elimination ●Quantitative analytical methods ●Sensitivity, characteristic concentration and detected limit ●brief introduction of atomic fluorescence spectrometry (AFS) 掌握原子吸收光谱分析法的基本原理,了解原子吸收光谱 仪的构造,重点掌握原子吸收分光光度定量分析方法。 Outlines
§8-1 Summarization of AAS ◼ 18th century, atomic absorption spectra was found ◼ 1955, AAS was first used on analytical chemistry by Walsh A. a Australian physicist
§8-1 Summarization of AAS ◼ 18th century, atomic absorption spectra was found ◼ 1955, AAS was first used on analytical chemistry by Walsh A. a Australian physicist
AAS Spectrophotometers: Readout E口 Shutter Amplifier Grating Ebert Modulated monochromator power source Lamp Flame (a)
原子吸收法的选择性高,干扰较少且易于克服。 由于原于的吸收线比发射线的数目少得多,这样谱线重 叠 的几率小得多。而且空心阴极灯一般并不发射那些 邻近波长的辐射线经,因此其它辐射线干扰较小。 原子吸收具有较高的灵敏度。 在原子吸收法的实验条件下,原子蒸气中基态 原于数 比激发态原子数多得多,所以测定的是大部分原 子。 原子吸收法 比发射法具有更佳的信噪比 这是由于激发态原子数的温度系数显著大于基态原子。 Comparison of AAS and AES 从仪器构造来看, 二者还有以下的异同点
原子吸收法的选择性高,干扰较少且易于克服。 由于原于的吸收线比发射线的数目少得多,这样谱线重 叠 的几率小得多。而且空心阴极灯一般并不发射那些 邻近波长的辐射线经,因此其它辐射线干扰较小。 原子吸收具有较高的灵敏度。 在原子吸收法的实验条件下,原子蒸气中基态 原于数 比激发态原子数多得多,所以测定的是大部分原 子。 原子吸收法 比发射法具有更佳的信噪比 这是由于激发态原子数的温度系数显著大于基态原子。 Comparison of AAS and AES 从仪器构造来看, 二者还有以下的异同点