GC工作过程 GC工作过程 Gas Chromatography(GC)is based on the ability of a chromatography column, containing the stationary phase,to separate the components of the sample as it is carried through the column by a flow of inert carrier gas.Only volatile samples can be analysed,and elevated temperatures are used to ensure adequate vapour pressures of each sample component.GC can be used to determine the number of components in a sample,and the concentration and purity of each component.The block diagram below shows the main stages of a GC analysis. Prepare Inject onto Separate Detect Identify and Sample Column Components Sample Quantify Components
一.GC工作过程
二.气路系统:包括气源、气体净化、气流控制和测量。 8 9 9 MAAA 图19-4补偿式双气路结构示意图 1一载气; 2一减压阀;3一净化器;4一稳压阀,5一压力表;6.6-针形阀 7.7-一转子流速计;8.8-气化室;9.9-色谱柱;10一检测器
二.气路系统:包括气源、气体净化、气流控制和测量
三.进样系统:包括进样器、气化室
三.进样系统:包括进样器、气化室
The split splitless injector Rubber septum +Septum purge outlet Carrier gas inlet Split outlet Heated metal block Glass liner Vapcurisction chamber Column
四.分离系统 分离系统由色谱柱组成,它是色谱仪的核心部件,其作用 是分离样品。色谱柱主要有两类:填充柱和毛细管柱。 1)植克往 填充柱由不锈钢或玻璃材料制成,内装固定相, 一般内径为2~4mm,长1~3m。填充柱的形状有U型和螺旋型 二种。 2)毛细馆柱 色谱动力学理论认为,气相色谱填充 柱在运行中存在严重的涡流扩散,影响柱效的提高。 1956年Golay由它的理论发明开管柱。它是一根内径 0.1-0.5mm,长度为10-300m的毛细管柱,每米理论 塔板数为2000-5000,总柱效最高可达10。 可分析复杂有机混合物,如石油成分、天然产物、 环境污染、生物样品
四.分离系统 分离系统由色谱柱组成,它是色谱仪的核心部件,其作用 是分离样品。色谱柱主要有两类:填充柱和毛细管柱。 1)填充柱 填充柱由不锈钢或玻璃材料制成,内装固定相, 一般内径为2~4 mm,长1~3m。填充柱的形状有U型和螺旋型 二种。 2)毛细管柱 色谱动力学理论认为,气相色谱填充 柱在运行中存在严重的涡流扩散,影响柱效的提高。 1956年Golay由它的理论发明开管柱。它是一根内径 0.1-0.5mm,长度为 10-300m的毛细管柱,每米理论 塔板数为2000-5000,总柱效最高可达106 。 可分析复杂有机混合物,如石油成分、天然产物、 环境污染、生物样品